Rigol DL3021A low current setting error

您所在的位置:网站首页 BiasREF Rigol DL3021A low current setting error

Rigol DL3021A low current setting error

2024-07-16 09:33| 来源: 网络整理| 查看: 265

When testing small currents, for example, currents less than 10 mA, my DL3021A does not apply the correct current until it reaches 8 or 9mA, then when dialing back to 0mA it seems to work until you reach 0mA then the trouble starts over.For example, setting the load to 0mA applying a source voltage 3.3V shows:

current setting    read value  (manually increasing by 1mA)=================0.000A               0.0021A0.001A               0.0022A0.002A               0.0023A0.003A               0.0023A0.004A               0.0023A0.005A               0.0024A0.006A               0.0024A0.007A               0.0025A0.008A               0.0025A0.009A               0.0095A 



【本文地址】


今日新闻


推荐新闻